3
III Installation
III.1 Installation
III.1.1 Unpacking and checking
1.
Gently remove the analyser and accessories from the packaging box. Save the
packaging materials for future transport or storage of the analyser. Check the
accessories against the packing list.
2.
Check the analyser and accessories to see if they are in good condition.
Notice:
If there are any problems, please contact Woodley Equipment Company.
III.1.2 Analyser placement
1)
The analyser should be placed in a clean and ventilated room with temperature
between 10°C ~ 30°C, relative humidity of less than 70%, away from direct sunlight.
2)
Make sure the vents are not obstructed and that there is at least 5cm of clearance
around the analyser.
3)
Connect the power adapter to the power interface of the analyser and turn on the
power.
4)
Do not place any items on top of the analyser.
III.2 Instructions
Please note that the operating temperature of test reagents is based on each test kit’s
instructions. Perform tests in strict accordance with the cartridge operating instructions
provided in each test kit.
III.3 Operation Procedures
III.3.1 Preparation
1)
When switched on, the analyser will run a self-test and the cartridge holder retracts
as shown below.
2)
The software will start automatically and display the main home screen.
3)
For the use and storage of reagents, please refer to the cartridge operating
instructions.
4)
Insert ID code chip for the test to be analysed. Select “Read ID Card”.
5)
Enter patient details in the ‘detail’ section.
2. Cartridge holder
1. Cartridge