3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
96
Figure 3-29 Replacement of the main control panel
1
2
1 - Main control panel
2 - M3×8 cross-recessed
pan-head combination screw (×4)
Installation
Follow the corresponding disassembly steps in reverse order.
Resetting the Machine
1. Assemble the back panel of housing.
2. Plug the power cord into the back panel of the device and turn on the power switch.
3.10 Reagent Testing Panel Replacement
Purpose
The reagent testing panel can be disassembled and replaced by following the procedures specified
in this section; however, in each case,
3.2 Preparatory Work before Repairs
before you begin.
Tools/Spare Parts
⚫
#2 (Ph2) Phillips screwdriver
⚫
The replacement reagent testing panel with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. Open the left side door.