2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
31
Name
Illustration
Usage
Remarks
plastic syringe
(disposal syringe
with no needle)
⚫
Used to manually unclog the WBC
and RBC channels
⚫
Used to inject other liquids.
None
Barrel connector
tubing connection
None
Silicone tube
Used to connect tubing and the plastic
syringe
None
Cotton swabs
⚫
Used to clean the swab
⚫
Used for other cleaning tasks
None
2.2.12.2 Inspection and Troubleshooting of Valve Clogging
1. To prevent the possibility of liquid flowing onto the base plate when the tube connected to the
valve is pulled out, the tube pulled out needs to be placed into the small sample cup, and tissue
paper should be used as a pad to prop up any affected components. The sample cup should be
removed once the liquid has stopped flowing.
2. Disassemble the valve. Then open the valve cap to check if there is any clogging caused by
impurities; if so, remove the impurities.
3. See Figure 2-20 and Figure 2-21.
Figure 2-20 3-way valve (L)
Figure 2-21 2-way valve (L)
4. Place the components in their original positions. (Note that the parts with instructions are kept on
the same side during installation. Then remove the tissue paper and clean the bottom.)