9. ADMINISTRATOR MODE
1.
Logical High
‐ Trigger activates if it is changed from Low to High level.
2.
Logical Low
‐ Trigger activates if it is changed from High to Low level.
You can see an example in the figure 9.9
(a) Logical High.
(b) Logical Low.
Figure 9.9: SMARTIS trigger level options ‐ logical High or logical Low with the Latch trigger example.
If you do not have any triggers available, you can control the SMARTIS via Ethernet commands. For this, check
the
Ethernet triggers
checkbox. You can find more information about the Ethernet commands in the chapter 11.
9.4.1 Frame Measurement and Evaluation
In this mode, you can start the measurement with one trigger. The measurement stops after defined sequence
length.
You can specify:
Sequence length
with number of frames to take.
Sequence length
in milliseconds.
Start Delay
‐ delay after the start‐trigger in milliseconds .
Figure 9.10: Frame Measurement and Evaluation example.