A310 Accupulser™ Pulse Generator
World Precision instruments
8
Output
AmPlitude ......................A ten-turn dial with an associated four-position rotary switch
are used to control the amplitude of the signal through
the -VAriABle and the +VAriABle output connectors. the
number set on the dial is the actual value of the output
voltage, depending on the full scale value selected by the
position of the AmPlitude switch. For example, 0.00 on
the dial and dc xV on the switch, represent a constant
-.0V on the -VAriABle and +.0V on the +VAriABle
outputs.
the dc positions represent a non-switching dc output
voltage even when the other outputs are providing
pulses. the Pulsed positions provide switched outputs,
synchronous with the other outputs. dc outputs with the
least amount of “noise” are obtained by switching the inPut
mode out of continuous on and not providing an
external input.
-VAriABle ...........................A Bnc connector used to provide a variable output signal.
signal polarity on the center pin is negative relative to the
connector shell.
+VAriABle .........................A Bnc connector used to provide a variable output signal.
signal polarity on the center pin is positive relative to the
connector shell.
ViolAtion .........................A red light, which is illuminated for approximately one
second, whenever the instrument is operating with timing
parameters that are in conflict with each other. Violations
are defined as follows:
eVent delAY is greater than eVent interVAl.
Pulse WidtH extends into the next eVent interVAl.
trAin durAtion extends into the next eVent interVAl.
Pulse WidtH is greater than Pulse interVAl.
Summary of Contents for A310 Accupulser