A310 Accupulser™ Pulse Generator
World Precision instruments
0
OPERATING DIRECTIONS
the model A30 Accupulser™ is designed to make the operation of the instrument
as self-explanatory as possible. extensive graphics are provided on the front panel to
guide the user with the function of each of the controls and connectors. Before using
the following instructions, it is recommended that the user should read about each
panel component in the controls, indicators and connectors section of this manual.
the following operating directions are provided as typical instrument setups, that
would result in valid operating modes.
Pulse
in the pulse mode, the instrument provides single or double pulses of user set widths.
the train generator is disabled by the user. in the single pulse mode, the pulse is
delayed after the start of the event interval. in the double pulse mode, a second pulse
is provided, starting at the same time as the event interval.
in the continuous on input mode, the time parameters for the event interval,
the event delay and the pulse width are set on the panel controls by the user. in the
eXt sYnc input mode, the internal clock is disabled and the event interval starts
whenever an input sync pulse is applied. in the sinGle eVent mode, one event
interval occurs for each push on the switch. in the eXt GAtes input mode, the panel
set event interval time is active as long as either of the gate signals is active. the eXt
GAtes feature allows the generation of a train of train pulses. At the outPut sYnc
connector, a narrow pulse is provided every time an event interval starts.
the output signal from the instrument is available at four different connectors. the
time parameters at each of the connectors are identical. the differences among the
outputs are due to the amplitude of the signal and the source/sink current capability
of each circuit. these differences are described in the specification section of this
manual.
it is possible to set the panel controls so that conflicting timing parameters are
provided. in order to assist the user, the ViolAtion light is illuminated for one
second, whenever one time interval improperly overlaps another. As a further aid, the
outPut light is illuminated for the duration of every output pulse and train duration.