A310 Accupulser™ Pulse Generator
World Precision instruments
5
MAINTENANCE
Cleaning
CAUTION: Do not use alcohol, aromatic hydrocarbons or chlorinated
solvents for cleaning. They may adversely react with plastic materials
used to manufacture the instrument.
the exterior of this instrument may be cleaned periodically to remove dust, grease
and other contamination. there is no need to clean the inside. use a soft cloth
dampened with a mild solution of detergent and water. do not use abrasive cleaners.
Test
the model
A310 Accupulser™
can be tested by following the steps in the operating
directions section of this manual. A different and more specific test is described
below. to perform all the tests, a voltmeter and an oscilloscope will be required.
set the front panel controls as follows:
inPut mode
continuous on
eVent interVAl
0 s
eVent delAY
550 µs
Pulse WidtH
00 s
sinGle/douBle Pulse sinGle Pulse
trAin GenerAtor
oFF
AmPlitude dial
0.00
AmPlitude switch
Pulsed x0V
Switch POWER to ON
Verify during the waveform “on” cycle, with the voltmeter, that the given voltages
exist at each connector.
-VAriABle
-0Vdc
+VAriABle
+0Vdc
monitor
+0 to +5Vdc
isolAtor
+3.5 to +5.Vdc