A310 Accupulser™ Pulse Generator
World Precision instruments
5
. locate the slide switch on the board and slide to the appropriate voltage setting.
the actuator is marked and the visible marking is the actual voltage setting, i.e. 5V
or 30V.
5. reassemble the instrument by reversing the order of the above steps.
Line Fuse
Warning: Before attempting fuse replacement, disconnect the instrument
from the ac power source.
the instrument contains one fuse, located in the fuseholder on the back panel. the
fuseholder is opened by inserting a screwdriver in the slot, pressing lightly, and
twisting the cap /8-turn counterclockwise. Verify that the fuse contained in the
fuseholder matches the desired line voltage. if necessary, replace the fuse with the
type and rating specified on the back panel.
Tilt Bail
the instrument is provided with a tilt bail. if use is desired, lift the front edge of the
instrument, swing the bail down and forward, and snap into place.
Rack Mounting
the instrument may be provided with a rack mounting kit for a standard 9 inch
rack. each mounting ear is assembled with two screws.