A310 Accupulser™ Pulse Generator
World Precision instruments
CONTROLS, INDICATORS & CONNECTORS
Front Panel
Input
mode ..................................A four position rotary switch selects the source of the
operation control signal.
continuous on ............provides a continuously running internal time base
established by the setting of the eVent interVAl controls.
eXt sYnc ...........................replaces the eVent interVAl time base by an external,
user-provided signal.
sinGle eVent ..................replaces the eVent interVAl time base by a one-time
signal provided by the user pressing the push button.
eXt GAtes .........................is the same as the continuous on mode, except that the
internal time base is only on when the appropriate signal is
applied to either of the two eXt GAtes connectors.
eXt sYnc ...........................Bnc connector input for an external signal to replace the
internal eVent interVAl time base. connector is only
active when selected by the mode switch. the signal should
be positive on the center pin, referenced to the shell.
sinGle eVent ..................A momentary push-button switch provides a single pulse
or sequence of pulses each time the switch is pressed. the
switch is only active when selected by the mode switch.
eXt GAtes .........................two Bnc connectors provide external signal input, which
switches the internal eVent interVAl time base on and off.
the connectors are only active when selected by the mode
switch. the VoltAGe signal should be positive on the center
pin, referenced to the shell. the sWitcH signal is a short-
circuit between the center pin and the shell.