16
*)
–
unit disabled for verified weighing devices.
Options in case when the main unit is [g]:
g (gram), kg (kilogram), ct (carat),
lb (pound)*.
*)
–
unit disabled for verified weighing devices.
Select start unit and press
key, next go back to the home screen,
to do it press
key.
Upon next start-up the scale runs with set start unit.
11.6.2. Temporary Unit
Temporary unit runs from the moment it is set to the scale shut-down and
restart.
Procedure:
Enter
<P9.Unit / 9.2.Unin>
submenu.
Press
key, available units are displayed successively one by one.
Options in case when the main unit is [kg]:
kg (kilogram), g (gram),
lb (pound)*, N (Newton).
*)
–
unit disabled for verified weighing devices.
Options in case when the main unit is [g]:
g (gram), kg (kilogram), ct (carat),
lb (pound)*.
*)
–
unit disabled for verified weighing devices.
Select temporary unit and press
key, next go back to the home
screen.
12. ADJUSTMENT
option available for non-verified scales exclusively
In order to ensure the highest weighing accuracy, it is recommended
to periodically introduce corrective factor of indications to scale memory, the
said factor must be referred to a reference weight. This is so called adjustment.