18
Procedure:
Enter
<P1.CAL / 1.2.CA-u>
submenu, edit box for declaring weight mass
is displayed (the mass value must be
30% of the maximum capacity
value).
Enter weight mass value and press
key to confirm, text
<UnLoAd>
(remove weight) is displayed.
Remove the load from the weighing pan and press
key.
Mass of an empty weighing pan is determined, this is signalled with
display of 'dash',
< - >
. Next, text
<LoAd>
(load weight) and mass
value that is to be loaded, e.g.
100 g
, are displayed.
Load the weighing pan with weight of specified mass value and press
key
.
Weight mass is determined, this is signalled with display of 'dash',
< - >
. Next, text
<UnLoAd>
(remove weight) is displayed.
Remove the load from the weighing pan,
<1.2.CA-u>
submenu
is displayed.
12.3. Adjustment Report
Adjustment report is automatically printed (using scale-connected printer) at the
end of each adjustment process. To declare report content go to
<P6.1.CrEP>
submenu. For detailed information concerning report content read later
sections of this manual.
13. SCALE PARAMETERS
Scale parameters are set to adjust the weighing device to ambient conditions
(filters) or individual needs (autozero on/off, tare values memory). These
parameters are to be found in
<P2.rEAd>
submenu.
<P2.rEAd>
submenu
comprises functions allowing you to adjust your weighing device to ambient
conditions of given workstation.
13.1. Filter
Enter
<P2.rEAd / 2.1.FiL>
submenu.
Press
key, filter values are displayed successively one by one:
1
- Fast,
2
- Average,
3
– Slow.