Communications and I/O
3 - 8
MedWeld 5000 Technical Reference Manual
M-032170
Caution: These inputs are arbitrarily assigned numbers
from 1 -8. Do NOT confuse this label (for example,
Input #1) used by the weld function with an actual input
position. That is, User Input #1 is NOT
NECESSARILY assigned to input #1.
Outputs
WELD VALVE Outputs (1 – 6)
These outputs are activated or de-activated when the MedWeld
3005 executes Function #54 (TURN ON WELD VALVE #n) or
Function #55 (TURN OFF WELD VALVE #n).
NOTE:
Valves #1 and #2 are shared outputs; they function as the two Advance
(Fluid) valves when enabling Ohma cylinder operation for the processor.
PRESSURE SELECT Outputs (1 – 4)
These outputs are activated or de-activated when the MedWeld
3005 executes functions #56 (TURN ON PRESSURE SELECT #n)
and #57 (TURN OFF PRESSURE SELECT #n).
RETRACT 1 VALVE
RETRACT 2 VALVE
These outputs respond to the on/off status of the RETRACT
VALVE 1 and 2 PILOT Inputs, based on the Retract Mode setup
parameter.
ISOLATION CONTACTOR Output
This output is activated or de-activated when the MedWeld 3005
executes Function #88 (TURN ON ISOLATION CONTACTOR) or
Function #53 (TURN OFF ISOLATION CONTACTOR).
Function #77 (ISOLATION CONTACTOR DELAY nn SEC) also
affects this output, as described in the setup parameters.
The output is used to remove power from the high-voltage circuitry
when it is not required by the weld schedule.
SHUNT TRIP Output
This output is provided to trip the circuit breaker in case of a
catastrophic failure.
Summary of Contents for MedWeld 5000
Page 2: ...MedWeld 5000 Technical Reference Manual Modified 1 31 06 M 032170...
Page 6: ...MedWeld 5000 Technical Reference Manual Modified 1 31 06...
Page 38: ...Installing the MedWeld 5000 2 12 MedWeld 5000 Technical Reference Manual M 032170...
Page 82: ...Advanced Software Features 5 12 MedWeld 5000 Technical Reference Manual M 032170...
Page 120: ...Hardware Troubleshooting 9 10 MedWeld 5000 Technical Reference Manual M 032170...