MedWeld 5000 Technical Reference Manual
5 - 1
M-032170
Advanced Software Features
5
This chapter describes some of the advanced features of the MedWeld
5000 control.
C-Factor
C-Factor is the value of current obtained for each change in percentage
of the maximum secondary current delivered. This value equals 1% of
the total available current during a weld.
The C-Factor can be used in various ways:
•
It can be used as a conversion factor between %I values and the
actual value of current that would be produced.
•
A change in C-Factor will indicate changes in the welding
environment.
The MedWeld 5000 calculates the actual C-Factor after every weld. The
programming device will display this value in its Weld Data display.
C-Factor is a theoretical maximum only. It does not take into account
the ratings of the weld control or transformer. C-Factor may possibly
exceed what the system can deliver, based on these ratings.
C-Factor is NOT accurate for extremely low currents, where the current
in the welding transformer’s secondary is not continuous.
C-Factor = I x x = x
pri
n
%I
I
%I
sec
V
nominal
V
line
V
nominal
V
line