Operation
105
Calibration
evaluation
After the calibration, the instrument automatically evaluates
the current status. The asymmetry and slope are separately
evaluated. The worst evaluation appears on the display.
Display
Asymmetry
[mV]
Slope [mV/
pH]
-15 ... +15
-60.5 ... -58
-20 ... +20
-58 ... -57
-25 ... +25
-61 ... -60.5
or
-57 ... -56
Clean the electrode
according to the electrode
operating manual
-30 ... +30
-62 ... -61
or
-56 ... -50
E3
Clear the fault according to
chapter 6 W
HAT
TO
DO
IF
...
< -30 or
> 30
< -62 or
> -50
1 Switch on the instrument by pressing
e
.
2 Connect the pH electrode to the instrument.
3 Keep the buffer solutions ready.
4 Adjust the temperature of the solutions and measure
the current temperature if the measurement is
performed without a temperature probe.
Preparatory
activities