Operation
112
8 Press the
g
key.
SLO(pe) appears on the display.
9 Thoroughly rinse the electrode with distilled water.
10 If necessary, set the temperature of the buffer solu-
tion using
u d
while keeping the
g
key de-
pressed.
11 Submerse the electrode in the second buffer solution.
12 Press the
g
key.
The second measured pH value appears on the dis-
play.
13 Set the nominal pH value of the second buffer solu-
tion (at the current temperature).
14 Press the
g
key.
The value of the slope (mV/pH) appears on the dis-
play.
The sensor symbol shows the evaluation of the elec-
trode after the two-point calibration.
15 Press the
g
key.
The value of the asymmetry (mV) appears on the dis-
play again.
16 To return to the measuring mode: Press the
m
key.