Operation
122
12 To return to the measuring mode: press the
m
key
or continue with the three point calibration.
13 Press the
g
key.
The previously selected concentration of the third
standard solution appears on the display.
14 Enter the current concentration of the third standard
solution using
u d
.
15 Thoroughly rinse the electrode with distilled water.
16 Immerse the electrode into the third standard solu-
tion.
17 Press the
g
key.
The electrode voltage appears on the display, the AR
display indicator flashes.
Three point
calibration