Operation
127
4.7.2 Switching on AutoStore (Int 1)
The storage interval (Int 1) determines the time interval
between automatic storage processes.
After the time interval expires, the current data record is
transmitted to the data storage and to the printer/interface.
The storage interval (Int 1) is set to OFF in the factory.
Thus, the AutoStore function is switched off.
To switch the function on, set up a time interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min).
Note
If AutoStore is switched on, the setting of the data transmis-
sion interval (Int 2) is ineffective (see see page 132).
1 Press and hold down the
g
key.
2 Press the
s
key. Int 1 appears on the display.
3 Press the
u d
keys to set up the required time in-
terval between storage processes.
4 Confirm with
g
.
The number of free memory locations appears on the
display.
Setting the storage
interval