Operation
141
4.10 Reset
You can reset (initialize) measuring and configuration pa-
rameters separately from one another.
Measuring parame-
ters
The following measuring parameters (pH InI) are reset to the
values they had on delivery:
Note
When the measuring parameters are reset, the calibration
data are lost. After the parameters have been reset, cali-
brate!
Configuration
parameters
The following configuration parameters (InI) are reset to the
values they had on delivery:
Measuring mode
pH
Asymmetry
of the pH electrode
0 mV
Slope of the pH electrode
-59.16 mV/pH
pH calibration procedure
AutoCal TEC
Temperature, manual
25°C
Resolution of
pH display
0.001
Baud rate
4800
Interval 1
(automatic storing)
OFF
Interval 2
(for data transmission)
OFF