What to do if...
150
Cause
Remedy
During measurement:
– Measured value is outside
the range determined by
calibration
– Select calibration
standards bracketing the
measured value
During calibration:
– Calibration standards are
too close together. Select
other calibration
standards
Cause
Remedy
During calibration:
– Calibration standards
have the wrong
temperature (maximum
temperature difference
±
2 °C)
– Temper the calibration
standards
During measurement:
– Test sample has the
wrong temperature
(maximum temperature
difference
±
2 °C)
– Temper the test sample
Cause
Remedy
– Standard 2 = Standard 1
– Use different calibration
standards
Concentration dis-
play flashes
U2
U1
–
5mV
≤
U3
U2
–
5mV
≤
TP flashes
Temp Error
Calibration
Std 2 flashes