What to do if...
151
6.3
General errors
You would like to
know which soft-
ware version is in
the instrument
Cause
Remedy
One of the following re-
quirements is not met:
– Standard 3 > standard 2,
if Std 2 > Std 1 or
– Standard 3 < standard 2,
if Std 2 < Std 1
– Use different standard 3
Cause
Remedy
– Batteries almost
depleted
– Replace batteries
(see section
5.1.1 C
HANGING
THE
BATTERIES
)
Cause
Remedy
– Timeout of the interface
– Checkout connected
instrument
Cause
Remedy
– Operating state
undefined or EMC
electric stress unallowed
– Processor reset:
Press the
a
key and
switch on instrument
Calibration
Std 3 flashes
LoBat display
to
to
to
to
display
Instrument does not
react to keystroke
Cause
Remedy
– e.g. question of the WTW
service department
– Press the
a
key and
switch on instrument.
The software version
is displayed.