pH 3310
Operation
47
ba75802d06
05/2017
mits the current dataset to the interface at the same time.
Configuring the
automatic storage
function
Settings
You can configure the automatic data storage function with the follow-
ing settings:
1
Press the
<STO
_
>
key.
The menu for automatic data storage appears.
Adjusted entire storage
duration
Max. available storage
duration
Graphical display of the
memory usage
ID number
1
Interval
30 s
Duration
180 min
Continue
0d03h00min
0
1d17h33min
Automatic data storage
15.03.2014
08:00
Back
Menu item
Possible setting
Explanation
ID number
1 ... 10000
ID number for the dataset
series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Storage interval.
The lower limit of the stor-
age interval can be
restricted by the number of
free storage locations.
The upper limit is restricted
by the storage duration.
Duration
1 min ... x min
Storing duration.
Specifies after which time
the automatic data storage
should be terminated.
The lower limit of the stor-
age duration is restricted
by the storage interval.
The upper limit is restricted
by the number of free stor-
age locations.