pH 3310
Operation
53
ba75802d06
05/2017
4.6.2
Connecting a PC
Connect the pH 3310 to the PC via the USB interface.
Attention
The USB interface is not galvanically isolated.
When a grounded PC is connected, measurements cannot be per-
formed in grounded media as incorrect values would result.
Installation of the USB
driver on the PC
System requirements of the PC for installation of the USB driver:
PC with USB port and CD-ROM drive
Microsoft Windows
(for details, see enclosed installation CD, Driver directory)
4.7
MultiLab Importer
With the aid of the MultiLab Importer software, you can record and eval-
uate measurement data with a PC.
1
Insert the supplied installation CD in the CD drive of your PC.
2
Install the driver from the CD.
Follow the Windows installation instructions as necessary.
3
Connect the pH 3310 to the PC via the USB interface.
The meter is listed as a virtual COM interface among the con-
nections in the Windows instrument manager.
More detailed information can be found in the MultiLab
Importer operating manual.