Operation
pH 3310
54
ba75802d06
05/2017
4.8
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.8.1
Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH
The following settings for pH measurements are reset to the default
settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<F1>
/[Menu] or
<ENTER>
key.
Setting
Default settings
Buffer
TEC
Cal. interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Resolution pH
0.001
Resolution mV
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
One point calibration
Off
Stability control
On
Temperature unit
°C