pH/ION 7320
Operation
ba75935e04
12/2017
25
5.4.6 Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date and time
are shown in the measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the fol-
lowing functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Therefore, check the time at regular intervals.
The date format can be switched from the display of day, month, year
(
dd.mm.yyyy
) to the display of month, day, year (
mm/dd/yyyy
or
mm.dd.yyyy
).
After a fall of the supply voltage (empty batteries), the date and time
are reset.
1.
In the measured value display:
Using
<MENU/ENTER__>
, open the
Storage & config
menu.
The instrument is in the setting mode.
2.
Select and confirm the
System
/
Clock
menu with
<
><
>
and
<MENU/ENTER>
.
The setting menu for the date and time opens up.
3.
Select and confirm the
Time
menu with
<
><
>
and
<MENU/
ENTER>
.
The hours are highlighted.
4.
Change and confirm the setting with
<
><
>
and
<MENU/ENTER>
.
The minutes are highlighted.
5.
Change and confirm the setting with
<
><
>
and
<MENU/ENTER>
.
The seconds are highlighted.
6.
Change and confirm the setting with
<
><
>
and
<MENU/ENTER>
.
The time is set.
Date format
:
dd.mm.yyyy
Date
:
12.12.2012
Time
:
12:12:25
Clock
12.12.2012
12:12
Back