pHotoFlex
®
pH
Operation
17
ba75977e03
10/2017
(polyurethane, ABS and PMMA). Thus, avoid contact with acetone and
similar detergents that contain solvents. Remove any splashes
immediately.
Cleaning the cell shaft
If liquid is in the cell shaft (e.g. due to a spilled cell), clean the cell shaft
as follows:
CAUTION
Cells can contain poisonous or corrosive substances. If the
content is released follow the danger warnings on the cell. If
necessary, take corresponding protective measures (protective
goggles, protective gloves etc.).
Cleaning the cells
Cells have to be clean, dry, and free of fingerprints and scratches.
Therefore, clean them regularly:
Scratches in the glass change the optical characteristics of the cell and
falsify the measured value. For this reason, never use scratched cells!
1
Switch the pHotoFlex
®
pH off and pull out the power plug.
2
Rinse the cell shaft with distilled water.
1
Clean the cells inside and out with hydrochloric acid or
laboratory soap.
2
Rinse out several times with distilled water.
3
Let them dry in the air.
4
Only hold the cells by the top or by the light protection cap so
that the optical path is not impaired.
5
Before measuring, clean the cell with the enclosed cleaning
cloth.