WUNDER SA.BI. SRL
Model: W1090 rev.00 20/10/2018
13
Indicator showing “0000“ ZERO SPAN out of range
1.
Incorrect weighing result - Avoid damages by external environment force such
as free-drop to the ground, collision by external objects, etc.
2.
Re-calibration procedure required to correct the setting of weighing accuracy.
3.
Interference due to RF disturbance, ground vibration…
4.
Unstable platform feet adjustments according to bubble level indication
5.
Incorrect position or other external objects within weighing area
6.
The scale is not in a solid and firm ground area, such as carpet floor or lawn.
Connection failure for data transmission to PC or printer
1.
Wrong connection wires or faulty wires for transmission between the digital
indicator and load cells.
2.
Wrong indicator model
3.
Wrong internal wiring or wire broken
WUNDER MODEL W1090 ELECTRONIC SCALE SERIAL N°..............................
We certify that this instrument has been inspected and has successfully passed
the functional test. It complies with the following standards and directives:
EN 45501 / EN60601-1-2 / EN60601-1
93/42/EEC – Medical Devices Directive
CONFORMITY
IDENTIFICATION LABELS
In the metrological label applied it’s displayed the year of manufactur ex. 17=2017 and so..