16
DANGER!
In case of voltages exceeding 30 V,
proceed with care as there is a risk of
electric shock.
Direct current measurement
■
Set the selector switch measurement range to
A
~
.
■
Set
M
button
[8]
to
DC
.
■
Connect the black measuring lead to the
COM
socket
[5]
and the red measuring lead to the
right socket (up to max. 250 mA).
Attention!
For measurements above 250 mA, the
10 A
socket
[7]
must be used.
■
Connect the measuring leads to the test object.
3
The multimeter automatically searches for the
most favourable measurement range.
■
Read the measurement result from the dis
-
play
[1]
.
Attention!
T
o protect the device against overheating, allow a
pause of 30 minutes for cooling after measuring
for a period of no more than 30 seconds.
Direct current
Measuring
range
Resolution
Accuracy
200 μA
0.1 μA
± 1.0 % of the mea
-
sured value + 3 digits
2 μA
1 μA
20 mA
0.01 mA
200 mA
0.1 mA
± 1.2 % of the mea
-
sured value + 5 digits
2
10 mA
10 A*
0.01 A
Overload protection: μA and mA range is protected
by F 250 mA / 690 V.
10 A range is protected by F 10 A / 690 V.
* Observe the maximum switch-on duration for the
10 A range!
Alternating current measurement
■
Set the selector switch measurement range to
A
~
.
■
Set
M
button
[8]
to
AC
.
■
Connect the black measuring lead to the
COM
socket
[5]
and the red measuring lead to the
right socket (up to max. 200 mA).
Attention!
For measurements above 250 mA, the
10 A
socket
[7]
must be used.
■
Connect the measuring leads to the test object.
3
The multimeter automatically searches for the
most favourable measurement range.
■
Read the measurement result from the dis
-
play
[1]
.
Attention!
T
o protect the device against overheating, allow a
pause of 30 minutes for cooling after measuring
for a period of no more than 30 seconds.
Alternating current
Measuring
range
Resolution
Accuracy
200 μA
0.1 μA
± 1.3 % of the mea
-
sured value + 3 digits
2 μA
1 μA
20 mA
0.01 mA
200 mA
0.1 mA
± 1.5 % of the mea
-
sured value + 5 digits
2
10 mA
10 A*
0.01 A
Overload protection: μA and mA range is protected
by F 250 mA / 690 V.
10 A range is protected by F 10 A / 690 V.
* Observe the maximum switch-on duration for the
10 A range!