X-Cite
®
TURBO: Quick Start Guide
SIDE 1
Warning
– Refer to Safety Precautions Booklet 035-00636R for all safety precautions prior to use.
User Guide 035-00601R download:
WARNING: Never look at operating LED! The light could severely damage the cornea and retina of the eye if the light is observed directly. Eye shielding
must be used at all times as well as clothing to protect exposed skin
Printed in Canada
Refer to user guide for more details.
035-00602R Rev. 2
SETUP
Unpack
1.
Carefully unpack the unit and accessories from shipping carton.
2.
When unpacking the light guide take care not to bend sharply.
Install the Microscope Adaptor (if applicable)
1.
Remove protective wrap from adaptor.
2.
Insert the flange portion of the Microscope Adaptor into the light port on
the microscope, and secure it using the hardware provided on the
microscope.
Connections and Positioning
1.
Remove red caps from light guide. Insert large end into light guide port on
unit (up to “B” line). Insert small end into microscope light guide port (or
adaptor), and secure it with hardware provided.
2.
Position the equipment. Ensure adequate clearance for ventilation around
unit, and slack in the light guide.
3.
Connect speedDIAL cable to “remote” port on main unit.
4.
Connect USB cable to ports on main unit and computer. (For Windows
7/8.1, USB driver will install via internet after unit powers up, see user
guide for additional information.)
5.
Connect external power supply (AC/DC switching adaptor) to main unit.
6.
Connect AC power to properly grounded outlet.
MANUAL OPERATION
1.
Flip rocker switch on rear of the main unit to turn X-Cite system on.
2.
When speedDIAL display shows LED channels 1 and 2 with their intensity
setting, it is ready to use.
3.
Press “up” and “down” buttons until the arrow is pointing to the desired
LED channel (1 to 6).
4.
Click speedDIAL to illuminate a specimen.
5.
Turn the dial to adjust intensity - clockwise to increase, counter-clockwise
to decrease.
REGULATORY (refer to user guide for complete details)
Product Safety and Electromagnetic Compatibility
The X-Cite TURBO has been tested and found to comply with product safety and
electromagnetic compatibility requirements. For a complete list of tests and for
certification details, please contact your X-Cite representative or visit Excelitas
website:
The X-Cite TURBO is classified as Class 1 laser product according to IEC 60825-1:
Safety of laser products – Part 1: Equipment classification and requirements: 2014-
05 and complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant
to Laser Notice No. 50, dated June 24, 2007. This class is eye-safe under all
operating conditions. This class includes all lasers or laser systems which cannot
emit levels of optical radiation above the exposure limits for the eye under any
exposure conditions inherent in the design of the laser product. There may be a
more hazardous laser embedded in the enclosure of a Class 1 product, but no
harmful radiation can escape the enclosure.
CE Marking
Council Directive 2014/35/EU
Low Voltage Directive
Council Directive 2014/30/EU
EMC Directive
Council Directive 2012/19/EU
WEEE Directive
Council Directive 2011/65/EU
RoHS
as amended by (EU) 2015/863
WEEE Directive
China RoHS RoHS
FCC
BASIC TROUBLESHOOTING
(refer to user guide for complete details)
Category
Symptom
Action
Error
messages
Error 1 (High Temp.)
Turn off, let cool; check
clearance around vents
Error 2 to 5 (Internal)
Turn off, restart
Error 6 (Low Temp.)
Allow unit to warm to room
temperature and restart
Intensity
Low output intensity
Check speedDIAL settings
Check light guide position
Check adaptor type
Check microscope beam path,
filters, apertures, shutters,
objectives
speedDIAL
New settings forgotten when
unit is shut down
After changing settings, wait at
least five minutes before
shutting unit down
Dial does not scroll through
the menu
Turn dial in the opposite
direction
“Lock” icon on display,
speedDIAL not responding to
manual control
Send “unlock” command via
computer OR power down unit
and restart