I N T E R N E G A T I V E O P E R A T I O N ( 8 8 2 O N L Y )
6-5
4. The display screen now indicates suggested changes for your filter pack.
These numbers are changes (indicated by the Greek letter
∆
) and should
be added to the filter pack used for exposing this internegative. Press key
l
(
∆
off) to turn the “change mode” off an view “absolute” filter pack
values.
∆
FP —03C +05M +02Y
∆
off
exp read
5. The absolute values are the addition of your starting filter pack (entered
earlier) and changed values. These values are adjusted to remove any
neutral density. Therefore, one of these values (usually cyan) is always
zero. These values constitute your new filter pack which should be used
for subsequent exposures. Press key
lll
(exp) to display the suggested
exposure.
FP 00C 39M 36Y
∆
on
exp read
NOTE:
Any neutral density adjustments to the filter pack affect exposure as
well. To avoid reciprocity problems, exposure time is held within ½ stop of
your starting exposure time. This sometimes causes an aperture change if the
calculated exposure time adjustment is greater than ½ stop.
6. Press
key
llll
(read).
EXP 11sec f/8.0
fp read
7. Measure another internegative. The entire procedure described is
repeated, except that the exposure and filter pack values just viewed are
used as the starting point.
Summary of Contents for 891
Page 1: ...891 892 A U T O M A T I C S T R I P R E A D I N G D E N S I T O M E T E R S Operation Manual...
Page 2: ......
Page 4: ......
Page 10: ...I N T R O D U C T I O N vi...
Page 34: ...S E C T I O N T H R E E 3 8...
Page 46: ...S E C T I O N F O U R 4 12...
Page 52: ...S E C T I O N F I V E 5 6...
Page 58: ...S E C T I O N S I X 6 6...
Page 60: ...S E C T I O N S E V E N 7 2...
Page 66: ...S E C T I O N E I G H T 8 6...
Page 68: ...A P P E N D I X A A 2...
Page 73: ...D 1 A P P E N D I X D Parts List and Packaging Drawings Parts List...
Page 74: ...A P P E N D I X D D 2 Packaging Drawing...
Page 78: ...A P P E N D I X E E 4...
Page 79: ......