RapidMatch
TM
GO Spectrophotometer
20
calibrate.
sample surface.
sure that all 3 sensor indicators
on the screen turn green.
Contact sensor(s) are
not working correctly.
Sensors may be stuck
due to dirt or paint; or
they are defective.
Open a new job to show the 3
sensor indicators on the screen
and place the instrument on a flat
surface and then lift it. If one or
more of the sensor indicators in
the screen do not change color
when lifted, there may be a
problem with the contact
sensor(s). Contact technical
support.
Shutter error.
Instrument defective.
Contact technical support.
Multiple error
messages.
Instrument encountered
a cascade of errors.
Press OK button in error
message for 2 seconds.
Date and time
wrong.
Incorrect setting.
Refer to instruction in manual.
16. Service Information
If you have any service request, please contact your local PPG Service and
Support.
17. Replacement Parts
Part
Part Number
USB cable
SE108-USBAB-03
Safety Strap
59-902288-01
Screen Protector
TPZ-27724