S E C T I O N T W O
2-2
CALIBRATING THE SCANNING INSTRUMENT
Calibrating the Scanning Instrument is virtually automatic, there is no
calibration reference to position. The Calibration References are a
permanent part of the Track assembly—located near the Docking Station.
The references are covered by a retractable cover when the Scanning Head
is away from the Docking Station, and exposed when the scanning head is
next to the docking station.
The Scanning Head automatically calibrates to the references when required
or when you request it.
NOTE:
The calibration references should be cleaned periodically to
maintain calibration accuracy. Refer to Section Three, General Cleaning for
procedure.
Retractable Cover