C i 7 X X 0 B E N C H T O P S P E C T R O P H O T O M E T E R
31
6.
When taking the transmission measurement the sample is held in position using the
transmission holder with a stop plate on the sphere side of the sample, and a spring loaded
clamping plate on the side of the sample away from the sphere.
7.
Attach the appropriate sample stop to the base plate and clamp to the carriage base with the
thumb screws. We recommend using a stop plate with all sizes including 25 mm.
8.
Remove the sample from the sample holder. Care should be taken to properly align the base
plate so that the sample is positioned between the stop and clamp toward the sphere. When
setting up for measurements, do not tighten the thumb screws until the sample has been
positioned properly.
9.
Once properly position, tighten the thumb screws and close the transmission cover.
10.
Initiate the measurement by selecting “Measure Standard" or "Measure Trial" from your
software application, or press the Standard or Trial button on the front panel. The
measurement is taken.
11.
Follow the instructions for saving the data in the application.