X-Rite DensiEye 100 and 700
DensiEye Reference Manual 4
25
4.2. Solid
reference
•
Reference values are referenced to paper white or absolute white base
•
Measured reference values appear in a table with solid density reference and
tolerance values for cyan, magenta, yellow and black
•
You can edit or measure these reference values and edit the tolerance values
for each color in this table
4.2.1. Measure Solid reference values (Paper)
•
Set White Base to Paper or Paper Auto in the Function Settings
•
Make sure that the instrument is calibrated for paper white base
•
Select Paper and measure paper white
•
After the measurement, the message box briefly displays “White Measurement
done”
•
Select
Solid
o
The output area displays a table with solid density reference and tolerance
values for cyan, magenta, yellow and black
o
Measure solid reference values
The output area displays measured solid density reference values
•
Select Return to return to the main menu
4.2.2. Edit Solid reference and tolerance values (Paper)
•
Set White Base to Paper or Paper Auto in the Function Settings
•
Make sure that the instrument is calibrated for paper white base
•
Select
Solid
o
Highlight solid density reference and tolerance values for cyan
o
Press the Enter button to open the edit or measure screen
•
Edit or measure solid density reference values
•
Edit solid density tolerance value
•
The output area displays the measured or edited solid density
reference and tolerance values
•
Continue editing the other colors as described
•
Select Return to return to the main menu