e X a c t A u t o - S c a n S y s t e m
17
Cleaning the System
Your system requires very little preventative maintenance to achieve years of reliable operation.
However, to protect your investment and maintain reading accuracy, a few simple-cleaning procedures
should be performed from time to time.
ATTENTION:
A properly grounded wrist strap is recommended when cleaning, removing or installing the
scanning head.
Make sure AC power is disconnected from the system before performing any instrument cleaning
procedure.
General eXact Auto-Scan Cleaning
The system covers may be wiped clean with a cloth dampened in water or mild cleaner.
DO NOT spray water or cleaning solution directly on the end covers or scanning track.
DO NOT use any solvents or harsh cleaners of any kind.
The track should be wiped clean with a lint-free cloth dampened in glass cleaner weekly or more often if
needed. When cleaning the track, make sure to clean the entire track. This includes the portion of the
track that resides under the eXact instrument trolley when it is in its docked position. You can slide it
over when cleaning is required.
System Contacts Cleaning
1.
Remove the eXact instrument from the scanning trolley and power it off.
2.
Turn the eXact instrument over and wipe the contacts on the charging adapter (1) weekly with a
cotton cloth dampened (not wet) in isopropyl alcohol.
3.
Make sure power is disconnected from the eXact Auto-Scan system.
4.
Wipe the contacts on the eXact charging station (2) weekly with a cotton cloth dampened (not wet)
in isopropyl alcohol.