e X a c t A u t o - S c a n S y s t e m
18
General eXact Cleaning
The exterior of the instrument and touch display should only be wiped clean with a cotton cloth
dampened (not wet) in water or isopropyl alcohol.
Cleaning the eXact Optics
The optics should be cleaned once a week in a normal environment. If the instrument is used in a dirty
or dusty environment, more frequent cleaning may be required.
1.
Remove the instrument from the scanning trolley.
2.
Rotate the reference holder (1) towards the back and hold.
3.
Blow short bursts of clean, dry air (2) into the measurement port. This should remove any
accumulated debris from the optics.
4.
Carefully return the calibration holder to its normal position.
IMPORTANT:
If can air is used for cleaning, do not invert or tilt the can during use. This could cause
damage to the optics.
Cleaning the eXact Calibration Plaque
The calibration plaque should be cleaned periodically.
Do not use solvents or cleaners of any kind.
The calibration tile is embedded in the underside of the reference holder.
1.
Remove the eXact instrument from the scanning trolley and open the target base.
2.
Rotate the reference holder (1) towards the back and hold.
3.
Blow short bursts of clean, dry air (2) onto the calibration tile. This should remove any accumulated
debris.
4.
Carefully return the calibration holder to its normal position.
(1)
(2)
(1)
(2)