e X a c t A u t o - S c a n S y s t e m
20
eXact Auto-Scan Specifications
Measurement System
Spectral Analyzer
DRS spectral engine
Spectral Range
400 nm – 700 nm with 10 nm interval
Measurement Geometry
45°/0° ring illumination optics, ISO 5-4:2009(E)
Light Source
Gas filled tungsten (illuminant type A) and UV LED
Measurement Conditions
M0, M1 (method 1 and method 2), M2, M3* (all according
to ISO 13655:2009)
*M3 option not available for Xp models
Calibration
On integrated white reference
Inter-instrument agreement
Average: 0.25
∆
E*ab, Max: 0.45 ΔE*ab (M3 0.55 ΔE*ab)
(Measurements using X-Rite manufacturing standards at a
temperature of 23°C +/- 1°C, 40-60% RH for all
measurement modes on 12 BCRA color tiles and a white
ceramic reference (D50, 2°))
Short term repeatability - White
0.05
∆
E*ab, (standard deviation) White BCRA
(Error compared to mean value of 10 measurements every
5 seconds)
Short term repeatability - Density +/-0.01 D for CMYK measurements
Data Interface
Bluetooth
Scan System
Track Lengths
29”/74 cm, 40”/102 cm
Min. Patch Size (Strip Height)
3.0 mm (with 1.5 mm aperture)
4.0 mm (with 2 mm aperture)
Scan Speed
1.5 mm aperture on 4 mm patches: >= 160 mm (7”)/sec.
Data Port
USB, Ethernet
eXact InkKeyControl
Software
Colors
Up to 16 colors
Print Standard Support
G7, PSO, Japan Color ISO 12647-2
Density Functions
Density, TVI (incl. SCTV), Trapping, Contrast, Gray Balance
Colorimetric Functions
CIE L*a*b*, CIE L*C*h°
dE methods
dE*76, dE*94, dE*00
Special Functions
BestMatch
Density White Base
Absolute, Paper
Pantone Color Libraries
Included
PantoneLIVE Support
Optional
Environmental
Operating Temp:
+10° (50°F) to +35°C (95°F)
Humidity Range:
0 - 85% RH non-condensing
Storage Temp:
-20°C to +50°C
Usage:
Indoor only
Altitude:
2000m
Pollution Degree:
2
Overvoltage:
Category II
Power Supply:
X-Rite P/N SE30-277, 100-240VAC 50/60Hz, 12VDC @ 2.5A
Vacuum Pump:
X-Rite P/N: SD07-07, 115/230 VAC 60/50 HZ 3.6/1.8 AMPS
Design and specifications subject to change without notice.