Connecting the Vacuum Pump Only (optional)
1.
Connect one end of the vacuum hose (1) to the barbed input fitting located on the vacuum pump,
and the other end to the barbed fitting on the right side of the system.
2.
Plug one end of the vacuum pump cable (2) into the connecter on the control box (3) and the other
end of the cable into the connector on the right side of the system.
IMPORTANT: Refer to the bottom of the page for the proper procedure to remove the vacuum
pump cable connectors from the eXact Auto-Scan and control box.
3.
Make sure the voltage selection switch (4) on the top of the control box (3) is set to the proper line
voltage for your region.
4.
Connect the detachable line cord (5) to the control box (3), and then plug the line cord into an easily
accessible, grounded AC wall receptacle.
Note: Only use appropriately approved AC Line Cord that
X-Rite provides. Any other AC Line cord can compromise the integrity of regulatory agencies
approvals for this product. Please contact X-Rite for replacement of the AC Line cord suitable
maintaining this products compliance
.
Proper Vacuum Pump Cable Removal Procedure
1.
Press downward on the clip at the top of the
cable connector.
2.
Slowly pull cable connector outward.
Vacuum hose (1)
Vacuum pump cable (2)
Control box (3)
Voltage selection
switch (4)
Line cord (5)
Vacuum pump cable (2)
Vacuum hose (1)
Connector clip