U
SER
M
ANUAL
X
ORAYA
Z7
|
X
ORAYA
ML-Z7
C
OMMISSIONING
31
Continuous current of the DC power supply
A 12 V DC power supply must deliver a continuous
current of 1 A with 3 A peaks. Use a regulated power
supply or a car battery and note the required voltage and
current values.
Please note the following sequence:
Connect the power cable to port
(L)
on the datalogger.
(
➔
Tighten the connector screws.
Connect the black plug of the cable to 0 V or ground.
Connect the red plug to the power supply.
The datalogger is securely connected to the measuring environment.
Once the supply voltage is established, the datalogger turns on and displays
its operational status via the power button LED
(K)
.
CAUTION
Damage due to improper device shutdown
Disconnecting the power supply during operation may
cause data loss and the destruction of the datalogger.
Never disconnect the datalogger from the power
supply while in operation.
Ensure proper connector seating and tighten the
screws if possible.
Only shut down the datalogger through the
X
ORAYA
Suite or the power button on the front panel.
Pull the black plug connected to ground last when
disconnecting the datalogger from the power supply.
Summary of Contents for XORAYA ML-Z7
Page 1: ...User Manual XORAYA Z7 XORAYA ML Z7 Edition 1 2 1 December 2022...
Page 72: ...USER MANUAL XORAYA Z7 XORAYA ML Z7 XORAYASUITE 72 RS 232...
Page 205: ......
Page 206: ......
Page 207: ......