U
SER
M
ANUAL
X
ORAYA
Z7
|
X
ORAYA
ML-Z7
C
OMMISSIONING
33
4.7
Configuring switches
Configure the built-in terminators via the rotary switches on the back
of the datalogger.
For the LIN interface, select between internal or external power supply
via the toggle switch.
4.7.1
4x CAN
Position
Resistance
0
Open (∞)
1
2.6 kΩ
2
120
Ω
The channels 3 and 4 have the value
Open
and, therefore, are not
terminated.
4.7.2
1x Dual-FlexRay
Position
Resistance
0
Open (∞)
1
2.6 kΩ
2
90 Ω
There are two variants of the label:
FR-2/1
and
FR-B/A
4.7.3
7x LIN
Position
Power supply
Vi
Internal
Ve
External
Use the internal power supply of the LIN measurement card only if the
connected LIN device uses the datalogger voltage as a reference voltage.
Bear in mind that this is not possible when using the datalogger in a 24-V
electrical system, for example. In this case, connect the external power
supply port to the corresponding reference voltage.