OPERation Status Register
M370078-01
5-37
5
Table 5-12
OPERation SHUTdown Status Register
Bit
Bit Weight
Bit Name
Description
0
1
PROTection
Reflects the summary of the PROTection sub-
register.
1
2
INTerlock
The power supply is shut down by INTerlock signal.
2
4
Not Used
Not Used
3
8
External Shutdown
The power supply is shut down by External
Shutdown signal.
Table 5-13
OPERation SHUTdown PROTection Status Register
Bit
Bit Weight
Bit Name
Description
0
1
OVP
Over Voltage protection has tripped.
1
2
UVP
Under Voltage protection has tripped.
2
4
OCP
Over Current protection has tripped.
3
8
Not Used
Not Used
4
16
AC Off
AC failure protection has tripped.
5
32
OTP
Over Temperature protection has tripped.
6
64
Not Used
Not Used
7
128
Foldback
Foldback protection has tripped.
8
256
Fan
Fan failure protection has tripped.