Local Operation
3-12
M370078-01
Constant Voltage Mode Operation
If the output is enabled and the configured current set point is much
higher than the requirements for the attached load, then the voltage will
rise until it reaches the voltage set point. When the output voltage reaches
the voltage set point, it stops rising. The unit will continue to operate in
CV mode. The load current will still vary to meet any changes in required
load current draw resulting from a change in the attached load.
Constant Current Mode Operation
If the output is enabled, the voltage and current start to rise. At this time
the unit is operating in CV mode (default operational mode). As the load
current drawn reaches the value of the output current setting, the unit will
switch to CC mode and not allow any additional current to be drawn
above the value of I
SET
. The unit is now operating in CC mode. The
voltage will still vary to meet any changes in required output voltage
resulting from a change in the attached load.
Automatic Mode Crossover
Mode crossover occurs when the unit makes the switch between operating
modes. Crossover can occur from CV to CC or vice versa, depending on
how the attached load resistance changes. Figure 3-4 shows the
relationships between the variables.
For example, if the unit was initially operating in CV mode with the
output voltage at the voltage set point and the load current below the I
SET
and the attached load resistance began to decrease, the current would rise
to meet this change in the load. The load current will increase in
proportion to the decrease in the load resistance until the load current
drawn reaches the current set point. At this point further decreases in load
resistance (increased load current requirements) will cause the unit to
cross over into CC mode where the output voltage will vary with
changing load resistance and the load current will remain at the I
SET
value.
The reverse operating mode change can also occur if the load resistance is
increased to the point that the required load current drops below the I
SET
value. At that point the power supply would crossover to CV mode and
the load current would be free to vary as the load resistance changed.
Summary of Contents for Watt Series Programmable DC Power Supply XG...
Page 2: ......
Page 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 30: ...1 8...
Page 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Page 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Page 198: ...5 72...
Page 262: ...A 36...
Page 284: ...IX 4...
Page 285: ......