SCPI Command Reference
IEEE-488.2/SCPI Syntax and Style
194
Operating Manual for XDC Series Power Supply
IEEE-488.2/SCPI Syntax and Style
Parameters
Units of Measure and Multipliers
Refer to
IEEE 488.2
, section 7.7.3 for the definition of units of measure.
The default units of measure include:
•
V (Volt – voltage)
•
A (Ampere – current)
•
W (Watt – power)
•
S (seconds – time)
The supported optional multipliers include:
•
m (milli)
•
k (kilo)
SCPI Command Hierarchy
SCPI is an ASCII-based command language designed for use in high-technology test
and measurement equipment. The command structure is organized around common
roots, or nodes, which are the building blocks of SCPI subsystems. An example of a
common root is CALibration, and some of the commands that reside in the
CALibration subsystem are shown below.
CALibration
:CURRent
[:DATA] <numeric value>
:LEVel {MIN|MAX}
[:SECure]
:CODE <new code>
:STATe {OFF|ON}, <code>
:STATe?
CALibration is the root keyword of the command. CURRent and SECure are
second-level keywords, and DATA, LEVel, CODE, and STATe, are third-level
keywords. A colon (:) is used to separate a command keyword from a lower-level
keyword.
Note
The SI standard for these multipliers is specifically lowercase, while the IEEE
standard specifies uppercase. Both combinations are supported.
Summary of Contents for XDC
Page 2: ......
Page 3: ...Operating Manual for XDC 6000 Watt and 12000 Watt Series Digital Programmable DC Power Supply ...
Page 22: ...List of Figures xx Operating Manual for XDC Series Power Supply ...
Page 180: ...Current Sharing 6000 Watt only Operation 178 Operating Manual for XDC Series Power Supply ...
Page 216: ...SCPI Command Reference Expressions 214 Operating Manual for XDC Series Power Supply ...
Page 228: ...GPIB Performance Specifications 226 Operating Manual for XDC Series Power Supply ...
Page 253: ......