25
Alarm output 'closed' when over-voltage trip occurs
26
Alarm output 'closed' when any trip occurs (factory default)
30 Buzzer
off
31
Buzzer on (factory default). A single beep indicates confirmation, a double
beep indicates a wrong entry.
40
Jog digit flashes, JOG indicator only flashes when jog is 'hidden'
41
JOG indicator always flashes, except when 'hidden' (factory default)
42
JOG indicator doesn't flash, except when 'hidden' (lazy flash)
91
Loads default calibration parameters. Refer to Service Manual
92
Shows firmware version number in the display
93
Sets these # settings to their factory default
99
Enter calibration mode. Refer to Service Manual.
Factory Default Settings
The ex-factory default settings (which will apply at first power-up) are as follows:
Range:
35V/3A (XDL 35-5); 56V/2A (XDL 56-4)
Voltage: 1·000V
Current Limit:
1·000A
OVP:
40V (XDL 35-5); 62V (XDL 56-4)
OCP:
5·5A (XDL 35-5); 4·4A (XDL 56-4)
Output:
Output off; local Sense
# Settings:
00
Output always off at power-up
26
Alarm output 'closed' when any trip condition occurs
31
Buzzer
on
41
JOG indicator always flashes; 'lazy' flash when hidden
RS232:
9,600 Baud (XDL 35-5P & XDL 56-4P only)
Address:
11 (XDL 35-5P & XDL 56-4P only)
Error Messages
The following hardware errors are indicated by showing the appropriate error number in the
display. The OK key will flash and if pressed the error will be ignored and operation will continue
as described.
Error No. Error Description
Action on pressing OK
1
Calibration constants corrupted at power-up
Loads default calibration parameters
2
# functions corrupted at power-up
Loads default # settings
3
Power-down settings not correctly loaded at
power up
Loads factory default power-up
settings
Switching the instrument off with the error message showing will leave all settings unchanged.
21