Looks Series (Surgical Telescope)                                                                                                      www.xenosys.co.kr 

 

10 

 

C.  The  telescope  provides  you  with  superior  optical  quality  and  lightness  for  all-day  wearing 

comfort. Head strap is equipped to distribute the weight to the nose & the head. Plug the end 

of temple arm into the end of head strap, and tighten the head strap. 

 

Picture of the head strap installation 

   

D. Loosen the head strap and take off the telescope. 

 

6. Precaution 

Please  keep  in  mind  that  you  will  have  a  hand-eye  coordination  adjustment  for  about  two 

weeks. 

A. Never damage, crush, or put a high pressure on the telescope, because the telescope is so 

sensitive and elaborate and this can cause other internal and potential damages.   

 

B.  Rinse  the  telescope  and  frame  with  the  supplied  cleaning  towel  or  soft  cloth  to  remove 

particles or debris and to prevent scratching. Use Phenol-Alcohol, Ammonia sterilizer or pure 

water with spray and  cloth.  Do not  use  disinfectants  containing  AUTOCLAVE,  CHEMICLAVE, 

GLUTARALDEHYDE,  IODOPHOR,  ACETONE  and  ultrasonic  cleaner  because  they  can  cause 

damage.   

C.  Use  carefully  because  scratching  and  contamination  can  cause  deteriorate  the  optical 

quality and function. 

D. Call the local sales  representatives  or  customer  center  if  you  have  difficulty  or  problem  by 

changes of working distance & eyesight. 

E.  Do  not  repair  your  telescope  by  unauthorized  technician  and  not  use  it  temporarily  when 

eyeglass lens or telescope lens is broken. It can cause headache and deteriorating eyesight 

by misalignment of pupil with telescope lens. Call the local sales representatives to repair our 

telescope 

F. Do not walk around with the telescopes on, be sure to take them off or flip up when walking 

around. 

Summary of Contents for Looks Series

Page 1: ...Looks Series Surgical Telescope User s Manual XENOSYS Co Ltd 2 Venture ro 100 beon gil Yeonsu gu Incheon 22013 Korea TEL 82 32 875 9811 2 FAX 82 32 875 9813 www xenosys co kr ...

Page 2: ... 1 Product Introduction 3 2 Product Description 3 3 Product Components 3 4 Adjustment 4 A Adjustment of TTL type before use 4 B Adjustment of Flip Up type before use 7 5 Operating Instruction 9 6 Precaution 10 7 Warranty 11 Appendex 1 TTL Flip Up Frame Figure 13 ...

Page 3: ...ts 1 Product Introduction The Looks series is medical equipment with Optic device to magnify the wound and is designed by mounting 2 surgical telescopes on eyeglass lens TTL type or frame Flip Up type TTL type has Looks2500 Looks2800 Looks3200 Looks3500 Looks5000 series and Flip up type includes Looks2500F series This combination manual consists of TTL type Flip up type instructions respectively 2...

Page 4: ...larity of object through telescope lens Check if the Working distance through telescope is similar to wanted working distance If any defect or damage is found contact the Sales representative or customer center 4 Adjustment A Adjustment of TTL type before use 1 Lever frame Check if your eyeglass rim are level with your eyes by adjusting the nose pad frame temples through a mirror to feel comfortab...

Page 5: ...ation angle and to make it clear Lift the frame to raise the axis of telescope lens for fitting if the upper image of field is blurred or vignetted through telescope If the lower image has such problem lower the frame and the axis When you lift the frame to raise the axis and to get clear image squeeze the nose pad or bend the end pieces of temple arms downward and recheck the image Left Image Rig...

Page 6: ...ge 4 Checking Pupilary Distance and Adjusting Check if the image through telescope is clear at your working distance If the image is blurred or vignetted through telescope as following figures adjust the frame leftward the axis of telescope lens the nose pad for fitting Left Image Right Image Nose pad Adjustment If the image is blurred or vignetted through telescope as following figures adjust the...

Page 7: ...telescope optics while you are adjusting the frame Don t not grab or attempt to turn the telescope or ocular B Adjustment of Flip Up type before use 1 Replacement of Eyeglass lens If you use any eyeglass when observing the objects located at the distance corresponding to the Working Distance in case of the presbyopia myopia and or astigmatism then you must replace the dummy eyeglass lens already i...

Page 8: ...stance between the pupils Please turn the PD adjusting knob slowly until you see each ocular s field of view converging to one circle while looking at the one object located at the position corresponding to your Working Distance Incorrect Correct If several users are not involved if your telescopes is not adjusted for anyone else it is convenient to set locking screw for adjustment stability Pleas...

Page 9: ... the upper end part of the observed visual field through the telescope is unclear or spread the nose pad and or fit the temple arm upward if the lower end part is unclear Downward adjustment Upward adjustment 5 Operating Instruction A After adjustment take and use the surgical telescope Some professionals with telescope experience should use the telescope full time but most of first users could ha...

Page 10: ...eaning towel or soft cloth to remove particles or debris and to prevent scratching Use Phenol Alcohol Ammonia sterilizer or pure water with spray and cloth Do not use disinfectants containing AUTOCLAVE CHEMICLAVE GLUTARALDEHYDE IODOPHOR ACETONE and ultrasonic cleaner because they can cause damage C Use carefully because scratching and contamination can cause deteriorate the optical quality and fun...

Page 11: ...t to refuse to repair or replace the product if the product is deemed to be unsanitary G In no event shall the liability of Xenosys Co Ltd be greater than the original purchase price of the product H You shall store the warranty certificate carefully because it is not reissued I Purchaser shall have responsibility to return the product in appropriate packing materials so as to avoid accidental bre...

Page 12: ...epresentative EU Repräsentant 1 Company Cormedics Medizintechnik GmbH 2 Address Bahnhofstr 32 82041 Deisenhofen Germany 3 Tel Fon 49 0 89 666 58 79 15 4 Fax 49 0 89 54 999 030 5 E mail a huettl cormedics com 6 Web http www cormedics com K Outer package Label ...

Page 13: ...not twist or bend Nose pad Adjust for comfort Pad arm Adjust for balance Temple Adjust for better fitting Hinge Do not apply pressure while adjusting the frame End piece hold when bending hinge corners or adjusting the width of two temples 2 Flip Up Frame PD Knob Adjust Pupilary Distance Flip up Hinge Allow you to flip up your telescope for normal vision Glasses lens Temple PD knob Flip up hinge ...

Page 14: ...ooks Series Surgical Telescope www xenosys co kr 14 Glasses lens Replace with the lens having your eye correction Refer to Page 7 Replacement of Eyeglass lens Temple Adjust for additional better fitting ...

Reviews: