Product description
Product instruction diagram
This temperature recorder is mainly used for temperature monitoring and recording of food and medicine storage and transportation
environments, such as refrigerated bags, refrigerated trucks, containers, etc.
This product is connected to a computer via USB , and the recorded data is directly output as a PDF report; the temperature probe is
built-in, powered by a CR2450 wide temperature lithium battery, and the waterproof level reaches IP67 . The product is equipped with a
barcode.
Alarm type
:
Parameter description:
Record capacity:
16000 groups (Max), there is difference due to the recording cycle.
Delayed startup:
record after 1 or more minutes(custom) later after logger startup.
Alarm threshold:
the upper and lower temp limits of alarm can be set.
Alarm delay:
Alarm starts after over-temp duration exceeds 1 or more minutes(custom), instead of immediately.
H
Over upper temp limit
Below lower temp limit
L
Single type
Cumulative type
Single calculation of continuous over-temperature time
Cumulative calculation of all continuous over-temperature time
Operating instructions
Long press 5 sec, the green light flashes
5 times, logger starts normally.
Long press 5 sec, the red light flashes 5
times, logger stops normally.
Insert the logger into the USB
port of computer to view data
report directly or by software.
Tear off the barcode if
you need
Elog1/1N
Single Use Temperature Data Logger
Instruction Manual
Technical parameters
ll function-related parameters of this recorder are configured by the factory, and users can customize some of these parameters
according to their needs;
-30°C~70°C/-22°F~158°F
-30°C~70°C/-22°F~158°F
±0.5°C/±0.9°F
(
-20°C~40°C/-4°F~104°F
);
1°C/1.8°F (other)
Built-in NTC
capacity of 16000 groups (maximum)
6
、
15
、
30
、
60
、
90
、
120
、
180
days (customizable)
single type ( Sin ), cumulative type ( Cum )
Factory default (customizable)
USB2.0/NFC
PDF format data report
Built-in CR2450 wide
temperature lithium battery
2 years of storage and use in
normal temperature environment
IP67
93*58*8mm (length * section * height)
about 20g
Measuring temperature
Storage temperature
Temperature
measurement accuracy
Sensor
Recording
Recording period
Alarm type
Alarm threshold
Data interface
Report type
Power supply
Battery life
Protection grade
Dimensions
Weight
Note: Before the recorder is started, it is recommended to store at room temperature
NFC operation instructions (ONLY for Elog1N)
1. How to install APP?
·
Download by scanning the QR code or searching ”Xetron iCold” in app store.
·
Register a new account then sign in.
2. How to set parameters?
·
Connect to the logger:
Click the navigation button at the upper left corner
Select the NFC temp label (confirm the NFC function of the mobile phone is turned on)
·
Set parameters:
Click at the upper right corner to set parameters, keep close to logger then click ”OK” to save settings
(Note: parameters only can be configured before the logger starts)
3. How to view and export data?
·
Connect logger and phone app firstly as above mentioned. View data directly once connection is successful.
·
Click to select report format to export.
Front
*Please refer to the actual product
Back
Elog1/1N
Configuration Info
Waterproof
Plastic Bag
USB Port
LED Indicator
Button(Start/
Status/Mark/Stop)
Barcode Label
My Device
No Device
Enter Device Name or Model
NFC Temperature Tag
Place your phone above the Tag
APP
NFC
NFC Temperature Tag
EFM200000000
EFM200000000
°C
UTC+08:00
V1.2
Basic Settings
Logging Settings
Alarm Settings
Wait for be...
Max-
Min-
Configure Time
:
2021-08-26 14:30:45
Serial Number
Temperature Uint
Timezone
Firmware Version
Running Time
:
-- Data Points
:--
23.2°C
Name
HI
L2
Threshold
8.0°C
2.0°C
Delay
00h 30m 00s
00h 30m 00s
Type
Single
Single
Logging Interval
Start Delay
00h 00m 10s
00h 00m 00s
NFC Temperature Tag
EFM200000000
EFM200000000
°C
UTC+08:00
V1.2
Basic Settings
Logging Settings
Alarm Settings
Record
Max 23.4°C
Min 23.4°C
Started
:
2021-08-26 14:33:20
Serial Number
Temperature Uint
Timezone
Firmware Version
Running Time
:
00h 00m 00s Data Points
:
1
23.4°C
Name
HI
L2
Threshold
8.0°C
2.0°C
Delay
00h 30m 00s
00h 30m 00s
Type
Single
Single
Logging Interval
Start Delay
00h 00m 10s
00h 00m 00s
Start/Stop Info
Mark Info
Status Request
Paste Area