XPro - Operation Manual
16
black red
DC and AC Current Measurement
Use proper terminals, function, and range for the
measurement. When the testing leads are connected to
the current terminals, do not parallel them across any
circuit.
The DC Current ranges are: 2mA, 200mA, 20A
The AC Current ranges are: 2mA, 200mA and 20A
To measure DC or AC current, perform the following
steps:
1. Turn off power to the circuit. Discharge all high-
voltage capacitors.
!
WARNING:
Never attempt an in-circuit current
measurement where the open circuit voltage
between terminals and ground is greater than
250V. If the fuse fails during measurement, the
Meter may be damaged or the operator may be
injured.