background image

XOS  

|

  801377-02  

|

  Rev D  

|

  English

 

 

 

 

 

Page 

3

 of 

22 

 

 

Electrical Exposure

: If any main system 

components are opened or dissembled, 

electrical components will be exposed that can 

produce a dangerous electrical shock. The X-

Ray components operate at high electrical 

voltages of up to 50 kilovolts, which can be 

lethal. It is the customer’s responsibility to 

ensure the electronics are properly grounded 

to earth. Service of the analyzer must be 

performed by authorized service personnel 

who are trained to work with high-voltage 

equipment. 

 

Thermal Exposure

: The components inside 

the equipment can reach high temperatures 

that can be potentially hazardous to the user. 

 

WARNING

 

This product can expose you to chemicals 

including lead/lead compounds which are 

known to the State of California to cause 

cancer or birth defects or other reproductive 

harm.  For more information, go to: 

            

www.p65warnings.ca.gov

 

 

Radiation Safety Statement 

This Product conforms to all applicable Standards under U.S. 

21 C.F.R. Chapter I, Subchapter J and IEC EN 61010-1. 

Each product is delivered with a radiation survey report that 

summarizes the radiation emission testing performed at the 

XOS factory before shipment.  The Product should be 

surveyed periodically by appropriate radiation safety personnel 

or qualified outside contractors.  

 

Radiation Safety Compliance 

This Product is sold with the understanding that the user 

assumes sole responsibility for radiation safety, as well as any 

state, provincial, or local regulatory compliance.  Such 

regulations may require but are not limited to: machine 

registration, personnel qualification, periodic survey 

requirements, maintaining a safety program, and/or 

maintaining a dosimetry program.  Users must consult the 

applicable regulating bodies for further information on user 

responsibilities.  Consult the full User Manual shipped with the 

Product for detailed Product description and additional safety 

and radiation information

.  

 

Summary of Contents for Petra 4294

Page 1: ...801377 02 REV D better analysis counts Basic User Manual...

Page 2: ...uide Used in any manner not specified by XOS Operated in an unsuitable environment Modified repaired or serviced by anyone other than XOS authorized personnel Subject to an accident mishap disaster or...

Page 3: ...BY QUALIFIED PERSONNEL Operating it as described in this guide and the full User Manual ensures that radiation safeguards work properly to limit radiation exposure keeping it well below the levels th...

Page 4: ...ww p65warnings ca gov Radiation Safety Statement This Product conforms to all applicable Standards under U S 21 C F R Chapter I Subchapter J and IEC EN 61010 1 Each product is delivered with a radiati...

Page 5: ...nge Sulfur 2 6 ppm 10 wt Limit of Detection Sulfur 2 6 ppm at 600s Applications Sulfur in hydrocarbons like crude oil diesel gasoline jet fuel and lubricants Product Specifications Method Compliance A...

Page 6: ...XOS 801377 02 Rev D English Page 5 of 22 Line Drawing Front View Power Button Screen Drip Tray Radiation Warning Light Sample Lid USB Port...

Page 7: ...XOS 801377 02 Rev D English Page 6 of 22 Line Drawing Back View Ethernet Port Power Cable Connector USB Port Serial Number Label Sample Lid...

Page 8: ...f 4 x 10mL 1 Power Supply 160W 100 240V AC input 24V DC output Quantity Description 1 XOS USB containing Petra documentation 8 Petra documentation 1 Radiation Safety Test Report 2 Petra Printer Docume...

Page 9: ...n and can start using the analyzer Setting the Date Time Data files created during measurements are coded with the date and time the measurement was initiated It is critical that the date and time of...

Page 10: ...g this action and informing the user that the unit will restart itself f After the unit reboots itself notice the change in the date and time Optional Device Setup To optimize usage you can connect th...

Page 11: ...urement Shutting Down the Instrument To shut down unit 1 Press the power button for 3 seconds or until the screen turns black 2 After about 30 seconds the power button should no longer illuminate mean...

Page 12: ...he Petra MAX XOS recommends the use of Etnom sample film This film can be reordered through XOS P N 200335 02 Mylar For the best results with the Petra 4294 XOS recommends the use of Mylar sample film...

Page 13: ...slide with the sample information 7 When the first measurement has successfully completed the unit will report the results a If the unit is still measuring the results will be reported in the History...

Page 14: ...a Preset 1 From the Scan Info screen select the button in the Preset section or follow the steps below 2 Find and select Settings located on the left side of screen 3 Locate and open Preset Manager 4...

Page 15: ...o make this Preset the default the preset selected when the analyzer initially starts up if desired g When done select Save Opening Results 1 From the home screen select the History button 2 If it is...

Page 16: ...scan file s HDX format is proprietary to Petra and is not compatible with any other software program Transfer to USB 1 Navigate to the Settings screen 2 Select the Transfer Data button 3 Insert an US...

Page 17: ...you want to remove the USB from the unit i If this is true then select OK ii If you want to copy more files select Cancel 7 Now you can remove the USB from the unit Transfer Data through the Network 1...

Page 18: ...ll be a security pop up The login is User name pi Password H ppy Gr0C33 7 When the Data folder is opened noticed that there will be date folders and a handful of other folders i e AppData Cal Config D...

Page 19: ...remove it from the sample interface and use cotton swabs with IPA If the basket is damaged replace it with a new basket Do not touch the sample window as this can affect your measurement results Prop...

Page 20: ...finement Calibration Set PetraMAX only 204715 01 USB powered hub For a more exhaustive list reference the Full User Manual xos com or call XOS Service Replacement Parts Before starting any maintenance...

Page 21: ...month If there is a spill or contaminants on the film it should be replaced immediately Replacing the Inlet Fan Filter 1 Gently pry the fan filter cover off the fan guard 2 Remove the filter 3 Lay the...

Page 22: ...til the message disappears which occurs when the x ray tube has cooled below 60 C The message pop up will NOT clear by pressing OK while the tube temperature is close to 65 C The message auto clear wh...

Page 23: ...rror IOP temp High The IOP is monitoring the source temperature which is too high Blocked vents fan failure or hot ambient temperature 1 Check for 6 inches of clearance around the analyzer 2 Check the...

Page 24: ...Contact the XOS Service Team Service 1 518 880 1525 service xos com xos com 15 Tech Valley Drive East Greenbush NY 12061...

Page 25: ......

Reviews: