System Specifications –
Electrical, Physical, and Operational
PetraMAX™ - Range, LOD & Applications
Dynamic Range
Sulfur 5.7 ppm - 10 wt.%
Limit(s) of
Detection
(ppm @ 600 s)
Sulfur 5.7 ppm
P
Cl
K
Ca
V
Cr
17
3
0.7 0.4
0.1
0.09
Mn
Fe
Co
Ni
Cu
Zn
0.07 0.07 0.07 0.04 0.1
0.1
Applications
Elements in liquid hydrocarbons like
crude oil, diesel, gasoline, jet fuel, and
lubricants. Also elements in catalyst
powder, petroleum coke / coal and
water.
Petra4294™ - Range, LOD & Applications
Dynamic Range
Sulfur 2.6 ppm - 10 wt.%
Limit of Detection
Sulfur 2.6 ppm at 600s
Applications
Sulfur in hydrocarbons like crude oil,
diesel, gasoline, jet fuel, and lubricants
Product Specifications
Method Compliance
ASTM D4294
ISO 8754
Measurement Time
30 - 900s
Calibration
30 calibration curves
Sample Film
PetraMax: Etnom®, 3.0 µm
Petra4294: Mylar, 3.6 µm
Sample Fill Volume
8 mL minimum
Data Output
Printout ● USB
Ethernet to PC connection
I/O Ports
Ethernet 10/100, USB
Power Supply
110-240 VAC ± 10%, 50-60 Hz
Operating Temperature
41°F to 104°F (5°C to 40°C)
Operating Humidity
30 – 85 %
Weight
28 lbs. (12.7 kg)
Dimensions (W x D x H)
14.5 x 16.5 x 6.0 in.
36.9 x 41.9 x 15.2 cm
Etnom® is a registered trademark of ChemPlex Industries, Inc.