12
Turn the Voltage control to adjust the OVP setting. When the desired setting is displayed press the output on/off button and the
supply will revert to its previous state.
To adjust OCP press and release the Current control. The following screen will appear:
Turn the Current control to adjust the OCP setting. When the desired setting is displayed press the output on/off button and the
supply will revert to its previous state.
If, during operation, any of these set points is exceeded the appropriate screen will be displayed as below.
Pressing Output On/Off when any of the trip screens are displayed will reset the trip and put the power supply back in idle
mode. The following screen will be displayed.
Pressing Enable at this time will turn on the output.
OcP set
OcP = 2.500
Overvoltage trip
Enable to reset
System set up
mode
Overcurrent trip
Enable to reset
Overpower trip
Enable to reset
Off
Ioc
V=100.0
i=3.500
7.2 Power Supply Setup
7.2.1 Introduction
Setup mode allows configuration of the power supply. Within setup one can adjust the power on state, select I/O options,
enable remote sense, and calibrate the power supply. To enter setup mode press and hold the output on/off button for at least 5
seconds. When enabled the initial setup screen will appear