background image

 

 

Data sheet 

 

Model 

XS BL 224 

Capacity 

220 g 

Precision level 

Division 

0.0001 g 

Repeatability 

± 0.0003 g 

Linearity 

± 0.0003 g 

Stable time 

≈ 3 sec

 

Warmup time 

60 

 120 minutes 

Operation temperature 

20 ± 2,5 °C  

Plate diameter 

90 mm  

Overall dimensions 

325mmX205mmX305mm

L*W*H

 

Chamber dimensions 

180mmX175mmX200mm

L*W*H

 

Net weight 

≈ 6kg

 

Adapter power supply 

460mmX360mmX400mm

L*W*H

 

 

Maintenance 

 

Cleaning 

 

Unplug the power adapter from the socket. 
If there is cable connecting with the balance interface, unplug the cable as well . 

 

Clean the balance by a cloth soaked in neutral flush (soap). 

 

After cleaning, wipe the balance by a soft dry cloth. 

 

Taking up the bracket to check the weighing system not damaged. 

 

Don’t make the liquid flow into the shield of balance; Don’t use the corrosive flush(solvent) to clean the 

stainless steel surface.  
All the stainless steel parts need to be flushed frequently, take out of all the stainless steel parts and flush 
them completely, use the wet cloth or sponge to clean the stainless steel parts.  
Only use the homely flush suitable to clean stainless steel parts.  

 
Warranty 

We offer two years warranty for the balance from the date it sold out. 
Damages not caused by misuse or vandalism all belong to warranty exclude the following case: 

 

Balance out of warranty. 

 

Damage caused by misuses such as long time under sunshine or receive corrosion. 

 

The user does not obey the operation manual such as overloading the balance. 

 

If the user disassembly the balance on his own, without authorized technician support. 

 

Any brute force application caused the damage of balance. E.g. throw the heavy loads to the pan, 
open or shut the chamber door violently. 

If you have any quality problem, please pack the product well together with the 
warranty card, then contact the distributor. 
The sensor is not within warranty range. 
Any product out of warranty or caused by misuse will be charged a reasonable repairing fee. 

Summary of Contents for BL 224 Basic

Page 1: ...1 BL 224 Basic User Manual...

Page 2: ...dangerous environment Please cut off the electricity without use for more than one week Please power off the balance when it connects with output device or before cut off the connection Magnet or stat...

Page 3: ...e use the balance Preparation before Use Adjust level before use Use the air bubble and feet under the front to adjust the level until the bubble move to the centre of the circle Put the balance on th...

Page 4: ...press 1d 10d conversion Normal weighing Preparation Connect the power supply Press On Off button to switch on the balance Pre heat Time In order to ensure the precision of weighing please preheat the...

Page 5: ...T Density measure mdoe DENSITY d Press MODE to confirm the selection Density measure mode is an option if you need please contact our company or authorized Distributor PCS counting By this procedure t...

Page 6: ...e solid density in air and then the density of the solid in a certain liquid the density of the liquid must be known in advance Procedure Display 1 Select Density Mode DENSITY d 2 Press CAL to confirm...

Page 7: ...BL xxx Backlight On Off AUTO BEPxxx Beeper On Off ModRec Function mode 0 switch on always in normal mode 1 switch on in the last weighing mode used Tadj Display temperature 1 9 to 1 9 Baud xx Baud rat...

Page 8: ...to be flushed frequently take out of all the stainless steel parts and flush them completely use the wet cloth or sponge to clean the stainless steel parts Only use the homely flush suitable to clean...

Page 9: ...be disposed of in public landfills According to the UE Directive 2002 96 EC the European users of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new...

Reviews: